Dynamic contact resistance (DCR) testing measures reed switch performance to prevent premature failures, improve reliability, and ensure fault-free operation in critical applications.

“Dynamic contact resistance is a “true” test and a way of measuring the contact resistance of reed switches to ensure fault-free operation, and maximum switching efficiency.”
If there are particular areas you need help with, feel free to skip to any of the following sections:
- Introduction
- What is a Dynamic Contact Resistance Test?
- What DCR Test Failures Look Like
- Dynamic Contact Resistance Test Factors
- Conclusion
- Additional Resources
Introduction
Reed switches simplify the process of energy transfer through a product or system, but there is much more to a reed switch than meets the untrained eye. The complexity of the act of “switching” itself and the number of factors that can impact the functionality of a reed switch are key concepts to understand. The health and function of a reed switch is a vital concern that has many threats, which is why Standex Detect recommends dynamic contact resistance testing.

What is a Dynamic Contact Resistance Test?
Dynamic contact resistance is a “true” test and a way of measuring the contact resistance of the reed switch blades. This testing ensures that all tools being utilised are not adversely affecting the reed switch. Furthermore, the test eliminates premature failures and improves long-term reliability in the equipment and technical systems.
Our typical dynamic contact resistance test consists of operating the reed contacts at about 100 times per second. At that rate, we look for a measurement of the contact resistance about 1 millisecond after the contacts close. If the contacts are clean (meaning no contaminants on the contacts) and the reed switch is intact, we usually get a positive result. However, if there is the slightest problem, the contact resistance will not have settled down within one millisecond timeframe, resulting in a rejected or failed reed switch. A DCR test is the most accurate and efficient way to examine the condition and capability of the reed switch. Additionally, DCR testing helps to determine if anything needs to be corrected during the manufacture of reed switches.

Many things can prove to be disruptive with regard to a reed switch’s function. The disruptions listed below can cause a DCR failure:
- Overstressed reed switch, usually from assembly
- Small crack on the reed seal
- Broken reed switch
- Plating or sputtering peeling off the contact area
- Air contamination in the glass capsule
- Particles on the reed contacts
What DCR Test Failures Look Like
Below are some examples of electrical patterns showing various DCR failures.
DCR Failure “Excessive Contact Bounce” indicating a weak contact closing force sure to shorten the life of the switch

DCR Failure “Excessive Dynamic Noise” and “Excessive Contact Wavering” both caused by a potentially stressed or cracked glass switch seal


DCR Failure “Changing Contact Resistance” caused by contact contamination, seal leak, particles, or peeling plating

Dynamic Contact Resistance Test Factors
When completing a DCR test, there are a few different factors that will determine the approach taken. For example, the test you run and the results you get will be based on the five factors below.
- Reed switch size
- The inductance of its coil
- Harmonic motion
- The voltage
- The overdrive level
Larger reed switches have more inertia and the reed blades are stiffer. This causes an increase in the initial reed closure time, the need for a magnetically stronger, more inductive coil, and an increase in the effects of the critically damped harmonic motion. Conversely, smaller reed switches have less inertia, are more flexible, and subsequently will behave in the opposite manner to larger reed switches. Reed switch size and the consequent inductance of its coil can have a major influence on the dynamic switching characteristics. When the reed contacts come together, they do so with a certain momentum that makes the reeds vibrate. This vibration is referred to as critically damped harmonic motion, which is an important concept in DCR testing.
When the reeds undergo critically damped harmonic motion, they are moving microscopically inside the glass capsule. The movement is occurring in the magnetic field generated by the coil, and when metal is in motion in a magnetic field, a current will be induced in the metal. This current is a critical part of the DCR test measurement. The overdrive of the coil is also a critical parameter in making the DCR measurement. Essentially, it is the voltage (or current) above the actual pull-in (or closure point) where the DCR measurement is made.
Conclusion
Reed switches provide us with a unique ability to utilise opening and closing contacts in virtually any environment with a simple magnet, enabling a large space for switching innovation. Dynamic contact resistance testing is a must in order to ensure fault-free operation, maximum efficiency and continued reed switching innovation. For this reason, Standex Detect DCR tests 100% of our reed-switch-based components to ensure its customers get the highest quality and reliability.
Get in touch with our team to learn more about our carefully engineered solutions.